Performance
Measuring Range
0.3 to 4.5 ppm
Number of Pump Strokes
4
Correction Factor
1
Sampling Time 2 minutes per pump stroke
Detecting Limit 0.2 ppm (n=4)
Colour Change White → Pale purple
Reaction Principle CHCl3 + I2O5 + H2S2O7 → Cl2
Cl2 + 3.3'-DIMETHYLNAPHTHIDINE ⅹ reaction product
Coefficient of Variation 10% (for 0.3 to 1.5 ppm), 5% (for 1.5 to 4.5 ppm)
Shelf Life 1 Year
Corrections for temperature & humidity Temperature correction is necessary
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Chlorine ℃6ppm + Pale purple
Bromine ″4.5ppm No No
Iodine ″4.5ppm No No
1,1,1-trichloroethane   + Pale purple
trichloroethylene   + Pale purple
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
10 ppm - -
仕様
目盛範囲: 0.3~4.5ppm〔基準n=4〕
測定範囲: 0.3~4.5ppm〔基準n=4〕
吸引回数〔n〕: 4〔基準〕
1回(100ml)の吸引時間: 2分
検知限度: 0.2ppm〔n=4〕
変色: 白色→淡紫色
温?湿度補正: 温度
有効期限: 1年(冷蔵庫保存)
指示精度: CV=5~10%
(CV:変動係数=σ:標準偏差÷平均値×100)
干渉ガス
ガス名 共存の場合の濃度と結果 単独の場合
塩素 2倍以上 +誤差 淡紫色に変色
1,1,1-トリクロロエタン   +誤差 淡紫色に変色
トリクロロエチレン   +誤差 淡紫色に変色
校正用ガス ガス拡散管法
許容濃度 管理濃度 TWA STEL 爆発範囲
3ppm 3ppm 10ppm -- --